Page 1 - Model 1054B DC

Model 1054B DCDual Cell Conductivity/Resistivity Microprocessor AnalyzerInstruction ManualPN 51-1054BDC/rev.BApril 2003

Page 2 - ESSENTIAL INSTRUCTIONS

4MODEL 1054B DC SECTION 1.0DESCRIPTION AND SPECIFICATIONS1054B DC REPLACEMENT PARTS AND ACCESSORIESPN ACCESSORIES2001492 Tag, Stainless Steel, Specify

Page 3

5MODEL 1054B DC SECTION 2.0INSTALLATIONSECTION 2.0INSTALLATION2.1 GENERAL. This analyzer's enclosure is suitablefor outdoor use. However, it shou

Page 4 - About This Document

6MODEL 1054B DC SECTION 2.0INSTALLATION2.4.1 Power Input Wiring. The Model 1054B can beconfigured for either 115 VAC or 230 VAC power.Connect AC power

Page 5 - MICROPROCESSOR ANALYZER

7MODEL 1054B DC SECTION 2.0INSTALLATIONFIGURE 2-1. Panel Mounting CutoutWHEN INCH AND METRIC DIMSARE GIVENMILLIMETERINCHDWG. NO. REV.41054B01 B

Page 6 - LIST OF TABLES

8MODEL 1054B DC SECTION 2.0INSTALLATIONFIGURE 2-2. Panel Mounting Tab InstallationDWG. NO. REV.41054A26 A

Page 7 - SECTION 1.0

9MODEL 1054B DC SECTION 2.0INSTALLATIONFIGURE 2-3. Pipe Mounting InstallationWHEN INCH AND METRIC DIMSARE GIVENMILLIMETERINCHDWG. NO. REV.41054B02 C

Page 8 - MODEL 1054B DC SECTION 1.0

10FIGURE 2-4. Electrical WiringMODEL 1054B DC SECTION 2.0INSTALLATIONDWG. NO. REV.41054B46 A

Page 9 - OPERATING RANGES

11FIGURE 2-5. Sensor WiringMODEL 1054B DC SECTION 2.0INSTALLATIONNOTE: TERMINALS IN JUNCTION BOX ARENOT NUMBERED.6OLD MODEL 140 SERIES SENSORS MAYNOT

Page 10 - 02 LED Display

12Configuration is all accomplished through a series ofmenus located within the set mode menu. To accessthese set mode menus the ACCESS keypad ispres

Page 11 - INSTALLATION

13MODEL 1054B DC SECTION 3.0DESCRIPTION OF CONTROLSFIGURE 3-4. LCD DisplayRELAY 1ACTIVATEDRELAY 3ACTIVATED(MINUS SIGN)RELAY 2ACTIVATEDCELL SELEC-TION:

Page 12 - MODEL 1054B DC SECTION 2.0

WARNINGELECTRICALSHOCK HAZARDMaking cable connections to and servicing this instru-ment require access to shock hazard level voltageswhich can cause d

Page 13 - 41054B01 B

14TABLE 3-1. Key DescriptionDisplays - current output (mA or % full scale).Set Function (w/SELECT) - Simulates currentoutput.Displays - low current se

Page 14 - 41054A26 A

15MODEL 1054B DC SECTION 3.0DESCRIPTION OF CONTROLSMNEMONIC DESCRIPTION1°C Temperature °C Cell 12°C Temperature °C Cell 21°F Temperature °F Cell 12°F

Page 15 - 41054B02 C

16MODEL 1054B DC SECTION 3.0DESCRIPTION OF CONTROLS1Cin2Cin1CEL2CEL1FCt2FCt1tYP2tYPAL1AL2AL31t-C2t-C1OUt2OUtUErdFtdtSCOdCYCSEt1t2tOFFHiLoSP3H-LHY5dond

Page 16 - 41054B46 A

17MODEL 1054B DC SECTION 4.0CONFIGURATIONSECTION 4.0CONFIGURATION4.1 GENERAL. This section details all of the items avail-able in the Set Mode to conf

Page 17 - FIGURE 2-5. Sensor Wiring

18MODEL 1054B DC SECTION 4.0CONFIGURATIONTABLE 4-1.CONFIGURATION WORKSHEETUse this worksheet to assist in the configuration of the analyzer. Date: ___

Page 18 - DESCRIPTION OF CONTROLS

19D. Alarm Logic. Mnemonic H-L. Select this item forhigh or low alarm logic. High logic activates the alarmwhen the reading is greater than the set p

Page 19 - Í) Key. This key is

20MODEL 1054B DC SECTION 4.0CONFIGURATION5. If OFF is selected, display will show oFF to acknowl-edge. Press ENTER key to return to AL1 or AL2,conclud

Page 20 - MODEL 1054B DC SECTION 3.0

21MODEL 1054B DC SECTION 4.0CONFIGURATION9. SCROLL(Ï) to the desired item and ENTER it intomemory. Display will return to H-L. If changes to relayacti

Page 21

224.5.1 Current Output Configuration 1OUt / 2OUt. Refer toFigure 4-5.1. Enter Set Mode by pressing the ACCESS key twice.2. SCROLL (Ï) until 1OUt2OUt

Page 22 - FIGURE 3-5. Set Menu Items

232. Press the ALARM 1 or ALARM 2 key. SP1 or SP2will show briefly, followed by the Alarm 1 or Alarm2 Setpoint.NOTE:If the alarm is set to OFF or FLt

Page 24 - MODEL 1054B DC SECTION 4.0

24MODEL 1054B DC SECTION 4.0CONFIGURATION4.8 OUTPUT SCALE EXPANSION. The output is con-trolled as user configured. The output zero and full scalevalue

Page 25

254.9 SIMULATE CURRENT OUTPUTS. The outputscan be simulated to check the operation of devicessuch as valves, pumps, or recorders. The outputs canbe si

Page 26

264.10 DISPLAY CYCLE. Display Mnemonic CYC. Thisfeature allows the display to alternate between cellone and cell two PV readings. Refer to Figure 4-10

Page 27

27MODEL 1054B DC SECTION 5.0START-UP AND CALIBRATIONSECTION 5.0START-UP AND CALIBRATION5.1 START-UP AND CALIBRATION. Calibration andoperation of the M

Page 28

28NOTEAfter initial installation or any cleaning pro-cedure, the conductivity cell will require I to2 days after insertion into the ultrapurewater str

Page 29 - Set Menu

296.1 KEYBOARD SECURITY. Display Mnemonic COd.Select this feature to display the user defined securitycode. Any three digit number may be used for thi

Page 30

307.1 THEORY OF OPERATION. This section is a gen-eral description of how the analyzer operates. Thissection is for those users who desire a greater un

Page 31

31MODEL 1054B DC SECTION 8.0DIAGNOSTICS AND TROUBLESHOOTINGSECTION 8.0DIAGNOSTICS AND TROUBLESHOOTINGDisplay DescriptionEEP EEPROM write error (bad EE

Page 32

32MODEL 1054B DC SECTION 8.0DIAGNOSTICS AND TROUBLESHOOTING8.2 TROUBLESHOOTING. The Model 1054B analyzeris designed with the state of the art micropro

Page 33 - START-UP AND CALIBRATION

33MODEL 1054B DC SECTION 8.0DIAGNOSTICS AND TROUBLESHOOTING4. If the conductivity cell electrodes have becomeextremely dirty or fouled, chemical clean

Page 34 - Í) to display the con

About This DocumentThis manual contains instructions for installation and operation of the Model 1054BDCDual Cell Conductivity/Resistivity Microproces

Page 35 - KEYBOARD SECURITY

34MODEL 1054B DC SECTION 8.0DIAGNOSTICS AND TROUBLESHOOTINGTABLE 8-3. Troubleshooting GuideSYMPTOM PROBLEM ACTIONAnalyzer value not the same 1. Refere

Page 36 - THEORY OF OPERATION

35FIGURE 8-1. Simulate Conductivity InputMODEL 1054B DC SECTION 8.0DIAGNOSTICS AND TROUBLESHOOTING8.2.5 Instrument Electronic Check. This procedurewil

Page 37 - SECTION 8.0

36MODEL 1054B DC SECTION 9.0RETURN OF MATERIALSECTION 9.0RETURN OF MATERIAL9.1 GENERAL.To expedite the repair and return of instruments, propercommuni

Page 38 - MODEL 1054B DC SECTION 8.0

WARRANTYSeller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufacturedor Services pro

Page 39

Credit Cards for U.S. Purchases Only.The right people,the right answers,right now.ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITEhttp://www.raihome.com

Page 40

MODEL 1054B DC TABLE OF CONTENTSMODEL 1054B DC CONDUCTIVITYMICROPROCESSOR ANALYZERTABLE OF CONTENTSSection Title Page1.0 DESCRIPTION AND SPECIFICATION

Page 41

MODEL 1054B DC TABLE OF CONTENTSTABLE OF CONTENTS CONT'D.LIST OF FIGURESFigure No. Title Page2-1 Panel Mounting Cutout...

Page 42 - RETURN OF MATERIAL

1MODEL 1054B DC SECTION 1.0DESCRIPTION AND SPECIFICATIONSSECTION 1.0DESCRIPTION AND SPECIFICATIONS1.1 FEATURES AND APPLICATIONSThe Model 1054B Micropr

Page 43 - Irvine, CA 92606

MODEL 1054B DC SECTION 1.0DESCRIPTION AND SPECIFICATIONS21.2 PHYSICAL SPECIFICATIONS -GENERALEnclosure: Black ABS, with interior conductive coating,NE

Page 44 - The right people

3MODEL 1054B DC SECTION 1.0DESCRIPTION AND SPECIFICATIONS1.3 ANALYZER SPECIFICATIONS @ 25°CMeasurement Range: 0-20,000 µS/cm or 0-50.00 megohms-cmOutp